The Vibrio cholerae O1/O139 Combo Rapid Test Cassette (Feces) is a rapid chromatographic immunoassay for the qualitative detection of Vibrio cholerae O1 and Vibrio cholerae O139 in human feces, providing results in 10 minutes. The test utilizes antibodies specific for VC O1 and VC O139 antigens to selectively detect VC O1 antigens and VC O139 antigens in human feces.
¡¾DIRECTIONS FOR USE¡¿
Allow the test cassette, specimen, buffer and/or controls to reach room temperature (15-30°C) prior to testing.
For Fecal Specimens:
1. To collect fecal specimens:
Collect sufficient quantity of feces (1-2 ul or 1-2 g) in a clean, dry specimen collection container to obtain maximum antigens (if present). Best results will be obtained if the assay is performed within 6 hours after collection. Specimen collected may be stored for 3 days at 2-8°C if not tested within 6 hours.
For long term storage, specimens should be kept below -20°C.
2. To process fecal specimens:
For Solid Specimens:
Unscrew the cap of the specimen collection tube£¬then randomly stab the specimen collection applicator into the fecal specimen in at least 3 different sites to collect approximately 50 mg of
feces (equivalent to 1/4 of a pea). Do not scoop the fecal specimen.
• For Liquid Specimens:
Hold the dropper vertically, aspirate fecal specimens, and then transfer 2 drops of the liquid feces (approximately 80 μL) into the specimen collection tube containing the extraction buffer.
l Tighten the cap onto the specimen collection tube, and then shake the specimen collection tube vigorously to mix the specimen and the extraction buffer. Leave the tube alone for 2 minutes.
3. Bring the pouch to room temperature before opening it. Remove the test cassette from the foil pouch and use it within one hour. Best results will be obtained if the test is performed immediately after opening the foil pouch.
4. Hold the specimen collection tube upright and open the cap onto the specimen collection tube. Invert the specimen collection tube and transfer 2 full drops of the extracted specimen (approximately 80μL) to each specimen well (S) of the test cassette, then start the timer. Avoid trapping air bubbles in the specimen well (S). See illustration below.
5. Read results at 10 minutes after dispensing the specimen. Do not read results after 20 minutes.
Cat. No. |
Product Description |
Specimen |
Format |
Kit Size |
Cut-Off |
Status |
IVCAB-625 |
Vibrio cholerae O1/O139 Combo Rapid Test Cassette (Feces) |
Feces |
Cassette |
10 T |
See Insert |
CE |