¡¾DIRECTIONS FOR USE¡¿
Allow the test, specimen, buffer and/or controls to reach room temperature (15-30°C) prior to testing.
1. Remove the test cassette from the foil pouch and use it within one hour. Best results will be obtained if the test is performed immediately after opening the foil pouch.
2. Place the cassette on a clean and level surface.
For Serum or Plasma specimen:
· Hold the dropper vertically and transfer 1 full drop of serum or plasma (approximately 10 ul) to the specimen well, then add 1 drop of buffer (approximately 40 ul), and start the timer. See illustration below.
For Venipuncture Whole Blood specimen:
· Hold the dropper vertically and transfer 2 drops of whole blood (approximately 20 ul) to the
specimen area, then add 1 drop of buffer (approximately 40 ul ), and start the timer. See illustration below.
For Fingerstick Whole Blood specimen:
· To use a capillary tube: Fill the capillary tube and transfer approximately 20 ul of fingerstick
whole blood specimen to the specimen area of test cassette, then add 1 drop of buffer (approximately 40 ul) and start the timer. See illustration below.
3. Wait for the colored line(s) to appear. Read results at 15 minutes. Do not interpret the result after 20 minutes.
Note: It is suggested not to use the buffer, beyond 30 days after opening the vial.